Ultra-low Temperature Freezers
Lexicon
®
UniformityTestat-80°CCycle
TemperatureRecoveryAfter1MinuteDoorOpening
average uniformity of cycle = 4.78 °C
SUPERIOR SAMPLE PROTECTION
FAST PULL DOWN
EXCELLENT UNIFORMITY
Pull down time is the time required to bring
down the temperature of the ULT freezer from
ambient to the set temperature.
Esco Lexicon
® has a short pull down time
which signifies an efficient refrigeration
system*.
Esco Lexicon
® recovers temperature in minutes
following a 1 min. door opening*. Fast
recovery in temperature ensures integrity of
stored samples.
Uniformity test measures the difference
between the coldest spot and warmest spot
in the freezer chamber when the freezer is
operating at set temperature.
Esco Lexicon
® has excellent uniformity under
normal operating conditions*. This guarantees
that all samples are at the correct temperature.
1 101 201 301 401 501 601 701
-110
-120
-80
-90
-70
-50
-60
-100
FAST TEMPERATURE RECOVERY
AFTER DOOR OPENING
Temperature(°C) Temperature(°C)
Time(min)
Time(min)
8
-95
-100
-80
-85
-75
-65
-70
-90
1
8
15
22
29
36
43
50
57
64
71
78
85
92
99
106
113
120
127
134
141
148
155
162
169
176
183
190
197
204
211
218
225
232
239
246
253
260
267
274
281
288
295
302
309
316
323
Temperature(°C)
Time(min)
Chamber
Ambient
PullDownTestat20°CAmbient
1 201 401 601 801 1001
-80
-100
-20
-40
0
40
20
-60
Warm-uptimefrom-80°Cto-50°C**
Warm-up time measures the time taken for temperature to rise up
(from -80°C to -50°C) when the power is interrupted.
With a robust insulation system in place, Esco Lexicon® has the best
warm-up time among the freezers compared on the right*. Slower
warm-up during an extended power failure buys the user time for
carrying out a back-up plan.
EXTENDED WARM-UP TIME DURING
A POWER FAILURE
Warm-uptime(min)
SanyoMDF-
U73V(V.I.P)
Haier
DW-86L626
So-Low
U85-25
Thermo
Scientic
Revco
ULT2586-10-D
Thermo
Scientic
Forma906
120
1
1
0
190
130
210
190
110
120
304
UUS-668A-1